Systemic toxicity studies for medical devices according to ISO 10993-11
ISO 10993-11 systemic toxicity tests are performed as part of a medical device's biocompatibility evaluation to assess the risk of the device causing harmful effects throughout the body. Systemic toxicity refers to effects that may be observed when the device or its constituents enter the body and affect multiple organs or systems, rather than a specific contact site.
Systemic toxicity can be evaluated through studies of:
Acute systemic toxicity (exposure up to 24 hours)
Subacute systemic toxicity (exposure >24 hours up to 28 days)
Subchronic systemic toxicity (exposure for part of the lifespan)
Chronic systemic toxicity (exposure for a major part of the lifespan)
The appropriate study type is primarily determined by the device’s contact duration (cumulative duration is considered when the device is used multiple times).
Systemic toxicity tests are typically in vivo animal studies. Acute systemic toxicity studies are usually conducted on mice or rats, and the longer exposure studies on rats or rabbits. The device or its extracts are administered to the test animal via the most clinically relevant route (dermal, implantation, inhalation, intradermal, intramuscular, intraperitoneal, intravenous, oral, or subcutaneous administration). Exposed animals are then evaluated through clinical pathology, histopathology, and/or gross pathology. Samples are prepared in accordance with ISO 10993-12.
ISO 10993-1:2025 states that systemic toxicity can be evaluated through toxicological risk assessment (ISO 10993-17), which may help reduce animal testing. Implantation studies (ISO 10993-6) may also be combined with systemic toxicity testing – ask our experts for more information or a quote using the form below.
- Suitable sample matrices
- Medical devices (liquid, soluble, and solid)
- Required sample quantity
- Individually determined
- Available quality systems
- GLP test method
- Method expert
Business hours: Mon–Fri 9 AM – 5 PM Finnish time (EET/EEST)
Other tests we offer
Cleaning validation of medical devices according to ISO 17664
Irritation testing according to ISO 10993-23
ARE-Nrf2 luciferase assay according to OECD 442D - in vitro skin sensitization
Face mask tests according to EN 14683 for type II and type IIR masks
Biocompatibility evaluation of breathing gas pathways according to ISO 18562
Bacterial reverse mutation test (Ames) for genotoxicity evaluation
Pyrogenicity testing according to ISO 10993-11
Eye corrosion testing: OECD 492 EpiOcular assay
Neutral red uptake (NRU) assay for cytotoxicity assessment
Sterility tests according to ISO 11737-2: Aerobic bacteria, anaerobic bacteria and fungi
Ask for an offer
Fill in the form, and we'll reply in one business day.
Have questions or need help? Email us at [email protected] or call our sales team.